Zcal Commands
Zcal displays the stored and reference calibration information for each pulse width, then prompts you to enter a command.
The number of pulse widths may vary, depending on your system.
For a large number of pulse widths, use a large width terminal window.
All dates shown are in local time as configured on your computer.
----------- Horizontal Calibration -----------
Sto 1.0us Ref 1.0us
Fit Slope: 1.1838 1.1838 dB/dB
Cal: -34.51 -34.51 dBZ
Std Dev: 0.00 0.00
Noise Level: -93.11 -93.11 dBm
I0: -57.66 -57.66 dB
Constant: 68.84 68.84 dB
XMT Power: 200.00 200.00 KW
Burst Power: -10.10 -10.00 dBm
Dynamic Rng: nan nan dB
Siggen Low: -34.00 -34.00 dBm
Siggen High: nan nan dBm
IF at Low: -86.74 -86.74 dBm
IF at High: nan nan dBm
IF Maximum: -50.42 -50.42 dBm
Flag: 0 0
Cal Time: 12:27:21 12:27:21
Cal Date: 24 OCT 2002 24 OCT 2001
Siggen Date: 24 OCT 2012 24 OCT 2012
Siggen ID: 189-111 189-111
Current Siggen ID: 189-111
Current Siggen cal date: 24 OCT 2012
ZCAL command (? for help):
The following table lists the commands that you can enter at the prompt.
| Command | Description |
|---|---|
| Cal |
Enter a new calibration reflectivity |
| ID |
Set siggen ID |
| List |
Print out current numbers on terminal |
| Polar |
Switch between horizontal and vertical polarization |
| Quit |
Quit |
| Refer |
Set reference to Stored values |
| Write |
Write calibration file |
| ? |
Print help |