Dehydrator BITE Packets
The following tables list the BITE and Q-BITE packets for the LAB4.50 dehydrator and ETI ADH-2A dehydrator models.
Cibred LAB4.50 Dehydrator Packets
| Character | Function | Alarm/failure code | |
|---|---|---|---|
| 1 | SYNC (C0 Hex) | ||
| 2 | Identification byte (User Choice) | ||
| 3 | Dehydrator alarms | D0 = Low pressure alarm | S99 |
| D1 = High pressure alarm | S100 | ||
| D2 = High humidity alarm | S101 | ||
| D3 - D6 = Not used | |||
| 4 |
Failure code (model C16820) Not used (model C15130) |
D0 : Heater1 not working, S106 | |
| D1 : Heater2 not working, S107 | |||
| D2 : Pump1 not working, S108 | |||
| D3 : Pump2 not working, S109 | |||
| D4 : Humidity sensor - no signal, S110 | |||
| D5 : Communication failure between Ethernet module and dehydrator, S111 | |||
| D6: not used, S112 | |||
| 5 | Not used | ||
| 6 | D0 - D5 = Not used | ||
| D6 = Dehydrator not responding | S126 | ||
| 7 | D0 = Not used | ||
| D1 = Alarms present | S128 | ||
| D2 - D6 Not used | |||
| 8 | END OF MESSAGE (FF Hex) |
| Character | Function |
|---|---|
| 1 | SYNC (AF Hex) |
| 2 | Identification byte (User choice) |
| 3-4 | Pressure (hPa) (scale 1) |
| 5-6 | Flow rate (litres/hour) (scale 1) |
| 7-9 | Hour meter (scale 1) |
| 10-12 | Reserved |
| 13 | END OF MESSAGE (FF Hex) |
ETI ADH-2A Dehydrator Packets
| Char | Function | ||
|---|---|---|---|
| 1 | SYNC (C0 Hex) | ||
| 2 | Identification byte (User Choice) | ||
| 3 | Dehydrator status and fault conditions (Absorption Unit #1) | ||
| D6 = System Standby | S105 | ||
| D5 = System Leaky | S104 | ||
| D4 = AU #1 dead | S103 | ||
| D3 = AU #1 full | S102 | ||
| D2 = AU #1 regenerating | S101 | ||
| D1 = AU #1 in use | S100 | ||
| D0 = AU #1 idle | S99 | ||
| S1084 | Dehydrator status and fault conditions (Absorption Unit #2) | ||
| D6 = System Standby | S112 | ||
| D5 =System Leaky | S111 | ||
| D4 = AU #2 dead | S110 | ||
| D3 = AU #2 full | S109 | ||
| D2 = AU #2 regenerating | S108 | ||
| D1 = AU #2 in use | S107 | ||
| D0 = AU #2 idle | S106 | ||
| 5 | Dehydrator status and fault conditions (Alarms and warnings) | ||
| D6 = AU #1 won’t heat ALARM | S119 | ||
| D5 = Low Line Voltage warning | S118 | ||
| D4 = Low Temperature warning | S117 | ||
| D3 = High Temperature warning | S116 | ||
| D2 = Leaky System warning | S115 | ||
| D1 = High Pressure ALARM | S114 | ||
| D0 = Low pressure ALARM | S113 | ||
| 6 | Dehydrator status and fault conditions (Alarms and warnings) | ||
| D6 = Dehydrator not responding | S126 | ||
| D5 = Change Compressor warning | S125 | ||
| D4 = Dew point ALARM | S124 | ||
|
D3 = Leaky System / Excessive runtime ALARM |
S123 | ||
| D2 = AU #2 won’t cool ALARM | S122 | ||
| D1 = AU #2 won’t heat ALARM | S121 | ||
| D0 = AU #1 won’t cool ALARM | 120 | ||
| 7 | Dehydrator status and fault conditions (Alarms and warnings) | ||
| D6 = Spare | S133 | ||
| D5 = Spare | 132 | ||
| D4 = Spare | S131 | ||
| D3 = Spare | S130 | ||
| D2 = Units (1=metric, 0=English) | S129 | ||
| D1 = Alarms present | S128 | ||
| D0 = Warnings present | S127 | ||
| 8 | END OF MESSAGE (FF Hex) | ||
| Character | Function |
|---|---|
| 1 | SYNC (AF Hex) |
| 2 | Identification byte (User choice) |
| 3-4 | Pressure |
| 5-6 | Flow rate |
| 7-8 | Temperature |
| 9-10 | Duty cycle |
| 11 | END OF MESSAGE (FF Hex) |