AFC Functional Test (Optional)
Test Goal
Verify that the AFC properly tracks the burst pulse frequency.
Test Equipment
KVM connected
Test Procedure
- Enter the TTY setups through
dspx. - Enter the Ps mode and observe the output on a display scope.
- Verify the following:
-
Verify the system is in AFC mode by checking that the text on the terminal for the AFC % output says AFC.
-
Verify the frequency displayed on the setup terminal is within ±15 KHz of the center IF frequency (the default value for the AFC hysteresis outer limit in the Mb command); for example, in the range 29.985 ... 30.015 MHz.
If it is not in this range, verify that it moves within this range.
-
Turn radiate off for 10 minutes and then turn the radiate back on. Observe to see if the AFC properly tracks the magnetron frequency as the magnetron warms.
-
Set the control signal to the maximum and minimum values using MFC, then turn on AFC. Observe to see if the AFC properly tracks back to the correct frequency.
-
- Perform the tests listed above for each pulse width and verify that the AFC properly
tracks the center frequency:
- For pulse width 0.
- For pulse width 1.
- For pulse width 2.
- For pulse width 3.
Checklist
| Task | Checked OK/Not OK | Remarks |
|---|---|---|
| Test procedure completed successfully. | ||
|
Parameters are set. |
||
| Test Passed | ||
| For Customer | Date: | |
| For Vaisala | Date: | |